Sapphire™ infusion system
This unique patented technology, innovative design and intuitive touch screen, the Sapphire pump helps you deliver exceptional care. And, without hard keys to limit software updates, and a design with future functionality in mind, the Sapphire infusion system provides excellent opportunity for growth.
Demonstration Videos
Guidelines, Specifications & User Manuals
DETAILS
The Sapphire™ Infusion System, made by Hospira, is a unique unit that is built to last. Despite the Sapphire’s small, compact design, it is made with durable Lexan housing that is damage, dust and fluid resistant. The infusion system is designed with a full-color touchscreen, eliminating hard keys that could eliminate future software upgrades.
Reduce infusion preparation time with the Sapphire’s easily programmable preset options. The Sapphire also features quick priming capabilities and an easy snap-on, snap-off mounting system so you can spend less time configuring the unit and more time with your patient.
The Sapphire features multiple confirmation screens and authorization protocols to help maintain high safety standards through every stage of use. Seven alarm parameters address everything including possible occlusion errors, to air in the line and low battery. The Sapphire’s rechargeable, energy-efficient battery is rated for up to 24 hours of use.
Features
Six infusion modes: continuous, intermittent, TPN, PCA, multi-step and epidural
Touchscreen operations allow for unlimited upgrades, increasing the longevity of your device
Easily-programmable preset options reduce infusion prep time
Multiple authorization levels help ensure greater safety
Compact design fits into the palm of your hand
Lightweight unit weighs less than 15 ounces (without battery)
Durable Lexan housing resistant to damage, dust and fluids
Energy efficient, rechargeable battery rated for up to 24 hours of use
7 Safety alarms: air in line, cassette misplaced, occlusions, check for occlusion, downstream occlusion, upstream occlusion, flow error, insufficient battery
Detects both singular and accumulated air bubbles
Intuitive, full-color touchscreen is clear and easy to use
Cassette door not integral to pumping process
Extensive infusion history log catalogs up to 8,000 events
SPECIFICATIONS
Physical
Dimensions: 5.63” (14.3 cm) H x 3.78” (9.6 cm) W x 1.93” (4.9 cm) D
Weight: 14.7 (0.4 kg) oz without battery
Power
Battery life: 24 hours at 125 mL/hr
Power requirements: 100-240V, 50-60Hz, 0.6 A
Delivery
Volume range: 1 – 9999 mL (increments of 1 mL)
Delivery rate accuracy: ± 2.5% (subject to external conditions)
Delivery rates: 0.1 – 99.9 mL/hr with increments of 0.1 mL/hr; 100 – 999 mL/hr with increments of 1 mL/hr
KVO Rate: Up to 20 mL/hr with increments of 0.1 mL/hr
Alarms
System alarms: Air in Line, Cassette Misplaced, Occlusions, Check for Occlusion, Downstream Occlusion, Upstream Occlusion, Flow Error, Insufficient Battery
Occlusion Alarm Pressure: Up to 17.4 PSI (1.2 Bar)